|
Secondary
(Scatter) Radiation and Undercut Control
Secondary (Scatter) Radiation
Secondary
or scatter radiation must often be taken into consideration when
producing a radiograph. The scattered photons create a loss of
contrast and definition. Often secondary radiation is thought
of as radiation striking the film reflected from an object in
the immediate area, such as a wall, or from the table or floor
where the part is resting. Side scatter originates from walls,
or objects on the source side of the film. Control of side scatter
can be achieved by moving objects in the room away from the film,
moving the x-ray tube to the center of the vault, or placing a
collimator at the exit port, thus reducing the diverging radiation
surrounding the central beam.
It is often called backscatter when it comes from
objects behind the film. Industry codes and standards often require
that a lead letter "B" be placed on the back of the
cassette to verify the control of backscatter. If the letter
"B" shows as a "ghost" image on the film, a significant
amount of backscatter radiation is reaching the film. The image of the "B" is often very nondistinct as shown in the image to the right. The arrow points to the area of backscatter radiation from the lead "B" located on the back side of the film. The control of backscatter
radiation is achieved by backing the film in the cassette with
a sheet of lead that is at least 0.010 inch thick. It is a common practice
in industry to place a 0.005" lead screen in front and a 0.010" screen behind
the film.
Undercut
Another
condition that must often be controlled when producing a radiograph
is called undercut. Parts with holes, hollow areas, or abrupt
thickness changes are likely to suffer from undercut if controls
are not put in place. Undercut appears as a darkening of the radiograph
in the area of the thickness transition. This results in a loss
of resolution or blurring at the transition area. Undercut occurs
due to scattering within the film. At the edges of a part or areas
where the part transitions from thick to thin, the intensity of
the radiation reaching the film is much greater than in the thicker
areas of the part. The high level of radiation intensity reaching
the film results in a high level of scattering within the film.
It should also be noted that the faster the film speed, the more
undercut that is likely to occur. Scattering from within the walls
of the part also contributes to undercut, but research has
shown that scattering within the film is the primary cause. Masks
are used to control undercut. Sheets of lead cut to fill holes
or surround the part and metallic shot and liquid absorbers are
often used as masks.
|