|
Calibration
Methods

Calibration refers to the act of evaluating and adjusting the
precision and accuracy of measurement equipment. In ultrasonic
testing, several forms of calibration must occur. First, the electronics
of the equipment must be calibrated to ensure that they are performing
as designed. This operation is usually performed by the equipment
manufacturer and will not be discussed further in this material.
It is also usually necessary for the operator to perform a "user
calibration" of the equipment. This user calibration is necessary
because most ultrasonic equipment can be reconfigured for use
in a large variety of applications. The user must "calibrate"
the system, which includes the equipment settings, the transducer,
and the test setup, to validate that the desired level of precision
and accuracy are achieved. The term calibration standard is usually
only used when an absolute value is measured and in many cases,
the standards are traceable back to standards at the National
Institute for Standards and Technology.
In ultrasonic testing, there is also a need for reference standards.
Reference standards are used to establish a general level of consistency
in measurements and to help interpret and quantify the information
contained in the received signal. Reference standards are used
to validate that the equipment and the setup provide similar results
from one day to the next and that similar results are produced
by different systems. Reference standards also help the inspector
to estimate the size of flaws. In a pulse-echo type setup, signal
strength depends on both the size of the flaw and the distance
between the flaw and the transducer. The inspector can use a reference
standard with an artificially induced flaw of known size and at
approximately the same distance away for the transducer to produce
a signal. By comparing the signal from the reference standard
to that received from the actual flaw, the inspector can estimate
the flaw size.
This section will discuss some of the more common calibration
and reference specimen that are used in ultrasonic inspection.
Some of these specimens are shown in the figure above. Be aware
that there are other standards available and that specially designed
standards may be required for many applications. The information
provided here is intended to serve a general introduction to the
standards and not to be instruction on the proper use of the standards.
Introduction to the Common Standards
Calibration and reference standards for ultrasonic testing come
in many shapes and sizes. The type of standard used is dependent
on the NDE application and the form and shape of the object being
evaluated. The material of the reference standard should be the
same as the material being inspected and the artificially induced
flaw should closely resemble that of the actual flaw. This second
requirement is a major limitation of most standard reference samples.
Most use drilled holes and notches that do not closely represent
real flaws. In most cases the artificially induced defects in
reference standards are better reflectors of sound energy (due
to their flatter and smoother surfaces) and produce indications
that are larger than those that a similar sized flaw would produce.
Producing more "realistic" defects is cost prohibitive
in most cases and, therefore, the inspector can only make an estimate
of the flaw size. Computer programs that allow the inspector to
create computer simulated models of the part and flaw may one
day lessen this limitation.
The IIW Type Calibration Block

The standard shown in the above figure is commonly known in the
US as an IIW type reference block. IIW is an acronym for the International
Institute of Welding. It is referred to as an IIW "type"
reference block because it was patterned after the "true"
IIW block but does not conform to IIW requirements in IIS/IIW-23-59.
"True" IIW blocks are only made out of steel (to be
precise, killed, open hearth or electric furnace, low-carbon steel
in the normalized condition with a grain size of McQuaid-Ehn #8)
where IIW "type" blocks can be commercially obtained
in a selection of materials. The dimensions of "true"
IIW blocks are in metric units while IIW "type" blocks
usually have English units. IIW "type" blocks may also
include additional calibration and references features such as
notches, circular groves, and scales that are not specified by
IIW. There are two full-sized and a mini versions of the IIW type
blocks. The Mini version is about one-half the size of the full-sized
block and weighs only about one-fourth as much. The IIW type US-1
block was derived the basic "true" IIW block and is
shown below in the figure on the left. The IIW type US-2 block
was developed for US Air Force application and is shown below
in the center. The Mini version is shown on the right.
IIW Type US-1

IIW Type US-2

IIW Type Mini

IIW type blocks are used to calibrate instruments for both angle
beam and normal incident inspections. Some of their uses include
setting metal-distance and sensitivity settings, determining the
sound exit point and refracted angle of angle beam transducers,
and evaluating depth resolution of normal beam inspection setups.
Instructions on using the IIW type blocks can be found in the
annex of American Society for Testing and Materials Standard E164,
Standard Practice for Ultrasonic Contact Examination of Weldments.
The Miniature Angle-Beam or ROMPAS Calibration
Block

The miniature angle-beam is a calibration block that was designed
for the US Air Force for use in the field for instrument calibration.
The block is much smaller and lighter than the IIW block but performs
many of the same functions. The miniature angle-beam block can
be used to check the beam angle and exit point of the transducer.
The block can also be used to make metal-distance and sensitivity
calibrations for both angle and normal-beam inspection setups.
AWS Shear Wave Distance/Sensitivity Calibration
(DSC) Block

A block that closely resembles the miniature angle-beam block
and is used in a similar way is the DSC AWS Block. This block
is used to determine the beam exit point and refracted angle of
angle-beam transducers and to calibrate distance and set the sensitivity
for both normal and angle beam inspection setups. Instructions
on using the DSC block can be found in the annex of American Society
for Testing and Materials Standard E164, Standard Practice for
Ultrasonic Contact Examination of Weldments.
AWS Shear Wave Distance Calibration (DC)
Block

The DC AWS Block is a metal path distance and beam exit point
calibration standard that conforms to the requirements of the
American Welding Society (AWS) and the American Association of
State Highway and Transportation Officials (AASHTO). Instructions
on using the DC block can be found in the annex of American Society
for Testing and Materials Standard E164, Standard Practice for
Ultrasonic Contact Examination of Weldments.
AWS Resolution Calibration (RC) Block

The RC Block is used to determine the resolution of angle beam
transducers per the requirements of AWS and AASHTO. Engraved Index
markers are provided for 45, 60, and 70 degree refracted angle
beams.
30 FBH Resolution Reference Block

The 30 FBH resolution reference block is used to evaluate the
near-surface resolution and flaw size/depth sensitivity of a normal-beam
setup. The block contains number 3 (3/64"), 5 (5/64"),
and 8 (8/64") ASTM flat bottom holes at ten metal-distances
ranging from 0.050 inch (1.27 mm) to 1.250 inch (31.75 mm).
Miniature Resolution Block

The miniature resolution block is used to evaluate the near-surface
resolution and sensitivity of a normal-beam setup It can be used
to calibrate high-resolution thickness gages over the range of
0.015 inches (0.381 mm) to 0.125 inches (3.175 mm).
Step and Tapered Calibration Wedges

Step and tapered calibration wedges come in a large variety of
sizes and configurations. Step wedges are typically manufactured
with four or five steps but custom wedge can be obtained with
any number of steps. Tapered wedges have a constant taper over
the desired thickness range.
Distance/Sensitivity (DS) Block

The DS test block is a calibration standard used to check the
horizontal linearity and the dB accuracy per requirements of AWS
and AASHTO.
Distance/Area-Amplitude Blocks
-------
Distance/area amplitude correction blocks typically are purchased
as a ten-block set, as shown above. Aluminum sets are manufactured
per the requirements of ASTM E127 and steel sets per ASTM E428.
Sets can also be purchased in titanium. Each block contains a
single flat-bottomed, plugged hole. The hole sizes and metal path
distances are as follows:
- 3/64" at 3"
- 5/64" at 1/8", 1/4", 1/2", 3/4",
11/2", 3", and 6"
- 8/64" at 3" and 6"
Sets are commonly sold in 4340 Vacuum melt Steel, 7075-T6 Aluminum,
and Type 304 Corrosion Resistant Steel. Aluminum blocks are fabricated
per the requirements of ASTM E127, Standard Practice for Fabricating
and Checking Aluminum Alloy Ultrasonic Standard Reference Blocks.
Steel blocks are fabricated per the requirements of ASTM E428,
Standard Practice for Fabrication and Control of Steel Reference
Blocks Used in Ultrasonic Inspection.
Area-Amplitude Blocks
Area-amplitude blocks are also usually purchased in an eight-block
set and look very similar to Distance/Area-Amplitude Blocks. However,
area-amplitude blocks have a constant 3-inch metal path distance
and the hole sizes are varied from 1/64" to 8/64" in
1/64" steps. The blocks are used to determine the relationship
between flaw size and signal amplitude by comparing signal responses
for the different sized holes. Sets are commonly sold in 4340
Vacuum melt Steel, 7075-T6 Aluminum, and Type 304 Corrosion Resistant
Steel. Aluminum blocks are fabricated per the requirements of
ASTM E127, Standard Practice for Fabricating and Checking Aluminum
Alloy Ultrasonic Standard Reference Blocks. Steel blocks are fabricated
per the requirements of ASTM E428, Standard Practice for Fabrication
and Control of Steel Reference Blocks Used in Ultrasonic Inspection.
Distance-Amplitude #3, #5, #8 FBH Blocks
Distance-amplitude blocks also very similar to the distance/area-amplitude
blocks pictured above. Nineteen block sets with flat-bottom holes
of a single size and varying metal path distances are also commercially
available. Sets have either a #3 (3/64") FBH, a #5 (5/64")
FBH, or a #8 (8/64") FBH. The metal path distances are 1/16",
1/8", 1/4", 3/8", 1/2", 5/8", 3/4",
7/8", 1", 1-1/4", 1-3/4", 2-1/4", 2-3/4",
3-14", 3-3/4", 4-1/4", 4-3/4", 5-1/4",
and 5-3/4". The relationship between the metal path distance
and the signal amplitude is determined by comparing signals from
same size flaws at different depth. Sets are commonly sold in
4340 Vacuum melt Steel, 7075-T6 Aluminum, and Type 304 Corrosion
Resistant Steel. Aluminum blocks are fabricated per the requirements
of ASTM E127, Standard Practice for Fabricating and Checking Aluminum
Alloy Ultrasonic Standard Reference Blocks. Steel blocks are fabricated
per the requirements of ASTM E428, Standard Practice for Fabrication
and Control of Steel Reference Blocks Used in Ultrasonic Inspection.
|