|
-
Ultrasonic
Signal Interpretation
With the use of six easily fabricated blocks, students studying ultrasonic
testing can gain insight into transducer movement and indication behavior
on the ultrasonic flaw detector screen. The student should have prior
knowledge of screen range calibration and ultrasonic theory regarding
angle of incidence and reflection.
Equipment needed:
Ultrasonic Flaw Detector calibrated for a 2.5 inch screen range.
5 or 10 MHz transducer.
3 inch by 2.5 inch by 1 inch thick blocks.
The range of test blocks that can be used is nearly endless, but the
following are some suggestions.
Sample
1: The slot is cut parallel to top surface, and far enough from the
side of the block to allow a backwall reflect to be seen before the
transducer is scanned over the top of the slot. Student should see the
backwall reflection disappear and see multiple indications from the
slot. The slot is also a little less than 1/2T so when the echo from
the slot starts to appear the student should note a slight shift on
the time base line as the back echo loses amplitude and the indication
from the slot increases in amplitude.
Sample
2: A cavity is cut at an uneven angle to the top surface, and far enough
from the side of the block to allow a backwall reflect to be seen before
the transducer is scanned over the top of the cavity. Student should
see the backwall reflection disappear and see an indication from the
cavity appear. As the student scans left and right over the cavity,
they should note the indication on the screen move to the right of the
UT screen indicating an increase in depth from the reflector. Students
should also note the shape of the signal. Instead of a nice straight
up/down signal, the indication will have several small peaks included
with an indication that occupies a large section of the time base line.
Sample
3: This sample has 2 drilled holes, one close to the top and the other
close to the bottom. When the student scans over the top of the near
side hole, the signal may show up in the dead zone, dependent upon transducer
selection. If it does, have the student switch to a higher frequency
transducer and emphasize that resolution is better with higher frequency
and the dead zone is shortened. When scanning over the more distant
hole the student should note that the indication from the hole drops
off very quickly and the indication itself is a sharp indication that
does not occupy a large area of the time base line.
Sample
4: The slot is cut 0.15 inches below the surface with a drilled hole
underneath. As the student scans over the slot they will see multiple
back echoes coming from the slot. The first back echo may be in the
dead zone. Ask the student to study how the initial pulse changes when
they scan over the top of the slot. Question the student as to where
the drilled hole indication is on the screen. A lot of students do not
remember that you are unable to see anything under the first reflecting
surface. Have the student repeat the inspection from the opposite side
of the block and compare and contrast the two inspections.
Sample
5: This slot is cut at an angle of approximately 15 degrees. As students
scan over the top of the angled slot they will see loss of backecho
and a low amplitude indication from the slot. The indication will occupy
a large segment of the time base line. Have the student move the transducer
across the top of the slot from right to left and notice the movement
of the indication on the time base line. The indication moves to the
right on the screen as the slot goes deeper into the part and left on
the screen as the slot gets shallower in the part.
Sample
6: This block has a slot cut at an angle of approximately 30 degrees.
When the students scan the block they should see total loss of back
echo when over the slot. They may see an indication from the top edge
of the slot if this refection does not fall in the dead zone. They should
also see an indication that comes up beyond the location of the 1st
backwall echo. This indication is the result of the sound reflecting
from the slot, traveling to the upper left corner of the block and returning
to the transducer along the same path. Emphasize that the angle of incidence
is equal to the angle of reflection.

|